<< Click to Display Table of Contents >> Navigation: The System Automation Tab > Focus automation > Selecting and Configuring the Motor Controller > ErgoFocus |
NOTE: ErgoFocus will only work on Windows 10/11. Windows 7 or 8 will not install correct drivers!
The ErgoFocus controller communicates via a USB port (virtual serial port).
This is automatically identified (from the unique VID/PID).
|
When successfully connected, a green check mark is shown.
|
|
Steps/mm
The Steps/mm parameter must be set to correct Z scaling.
It is calculated as: motor[steps/rev] / Spindle[mm/rev].
Example: Stepper motor has 200[steps/rev], and spindle pitch is 0.25[mm/rev], this makes the Steps/mm parameter = 200 / 0.25 = 800.
Ramp Up/Down
This sets the start/stop acceleration ramp.
The bigger number, the slower acceleration. A value of 0 means 'instant' start/stop.
Swap Direction
Reverses the motor direction
Swap host direction
Reverses the sign of the position readout
Limit switch Normally Open
The polarity of the limit switches can be set. The default is "Normally Open".
Standby- and Running current
For ErgoFocus firmware ≥ v3.1.3 there is an option to adjust the standby- and running-current.
Adjusting the current requires great care and diligence.
DeltaPix can provide instructions on how to do this. Please contact [email protected].