Mitutoyo MF - Communication Setup

<< Click to Display Table of Contents >>

Navigation:  Appendix >

Mitutoyo MF - Communication Setup

 

Set up the microscope RS-232 communication to:

Before DeltaPix InSight v6.1.18 : 9600, N, 8, 1, NoFlowCtrl .

From DeltaPix InSight v6.1.18 : 9600, E, 7, 1 NoFlowCtrl (This is almost the default microscope setting)

 

 

See technical document 99MBA091A6 (31.05.2013) on how to configure the microscope.

See page 1-11 on how to setup RS232 communication

See page 1-13 on how to enter/exit setup

 

 

F07 Microscope Communication Setup

F07-1 = 9600 baud

F07-2 = 7 databits

F07-3 = Even parity

F07-4 = 1 stopbit

F07-5 = Flowcontrol (xon/xoff) = OFF

F07-6 = Flowcontrol (hw) = OFF

 

F03 Resolution setup

All axes, select the 0.1: Minimum readout of 0.1µm/0.00001”

 

 

RS232 Cable

clip0197

Optionally use a USB-RS232 converter in the PC end.

 

Dataformat example

 

Query : "A"<cr><lf>

Response : "X +004.1880,Y +004.5870,Z +006.1523"<cr><lf>

 

will display:

 

clip0237