<< Click to Display Table of Contents >> Navigation: Appendix > Installing and connecting to Zeiss MTB 2011 > Optics Editor and calibration hints |
(1) The MTBConfig of the objectives must be the same as in the Optics Editor!
Example of a AxioImager M2 configuration. The x5 is calibrated and interpolated to the other objectives.
(2) The calibration must be done with a 1x Tubelens and a 1x Reflector magnification!
Example of a valid calibration setup. Tubelens is 1x and Reflector is 1x.